Three Essential Elements of Medical Sealing Machines
Medical sealing machines used for final sterilization packaging must possess three essential elements: temperature, time, and pressure.
The sealing temperature varies depending on the material of the sterilized packaging bag, but the time and pressure remain relatively constant. For example, when using an MDcare sealing machine, the sealing temperature for high-pressure steam/ETO/formaldehyde sterilized paper-plastic packaging bags is 180℃;
For peracetic acid low-temperature plasma sterilized packaging bags (Tyvek bags), the sealing temperature is 120℃; and the sealing temperature for aluminum foil bags is mostly 80℃.
The sealing time is approximately 1.68 seconds. The pressure is between 80 and 120N, typically set at 100N. Note that different brands of packaging bags may have slight fluctuations in these heat and pressure parameters due to differences in the materials used.
Once these three elements of the sealing machine are set, they must remain stable during operation; for example, the MDcare sealing machine maintains a fluctuation of only 1%.
Recently Posted
-
Tres elementos esenciales de las máquinas de sellado médico
May 15, 2026Las máquinas de sellado médico utilizadas para el envasado de esterilización final deben contar con tres elementos esenciales: tem
Read More -
Função de Validação de Seladoras Médicas
May 14, 2026A norma final para embalagens de esterilização de dispositivos médicos, ISO 11607-2, exige que as seladoras médicas possuam funcio
Read More -
Três Elementos Essenciais das Seladoras Médicas
May 14, 2026As seladoras médicas utilizadas para embalagens esterilizadas devem possuir três elementos essenciais: temperatura, tempo e pressã
Read More -
Validation Function of Medical Sealing Machines
May 14, 2026The final medical device sterilization packaging standard ISO 11607-2 requires medical sealing machines to have validation functio
Read More